Theory and History:
Time-of-flight mass spectrometry (TOF-MS) is
probably the simplest method of mass measurement to conceptualise,
although there are hidden complexities when it comes to higher
resolution applications. The first commercial TOF instrument was
marketed by the Bendix corporation in the late 1950's. Their design was
based on the Wiley & MacLaren instrument that was published in 1955
[1]. TOF-MS has really come into its own in recent years as being an
essential instrument for biological analysis applications - this is
especially the case with the coupling of TOF-MS to MALDI and ESI
ionisation methods and the development of high-resolution and hybrid
instruments (for example Q-TOF and TOF-TOF configurations). The
inherent characteristics of TOF-MS are extreme sensitivity (all ions
are detected), almost unlimited mass range and speed of analysis
(modern instruments can obtain full spectra in seconds). This makes
TOF-MS one of the most desirable methods of mass analysis.
Fig. 1: A Schematic of a
Time-of-Flight mass spectrometer operating in Reflectron Mode
The general set up of TOF is shown in Fig. 1. The
ions are introduced either directly from the source of the instrument
or from a previous analyser (in the case of Q-TOF) as a pulse. This
results in all the ions receiving the same initial kinetic energy. As
they then pass along the field free drift zone, they are separated by
their masses, lighter ions travel faster. This enables the instrument
to record all ions as they arrive at the detector and so accounts for
the techniques high sensitivity. The equation governing TOF separation
is:

|
m/z
is mass-to-charge ratio of the ion
E is the extraction pulse potential
s is the length of flight tube over which E is applied
d is the length of field free drift zone
t is the measured time-of-flight of the ion
|
Theoretically then, all the ions are given the same initial kinetic
energy by the extraction pulse and then drift along the field free
drift zone where they will be separated so that all ions of the same
m/z arrive at the detector at the
same time. In practice, the pulse is not felt by all ions to the same
intensity and so a kinetic energy distribution for each discrete
m/z exists. This lowers the
resolution by creating a time-of-flight distribution for each
m/z [2]. This is relatively easily
corrected for by the application of a reflectron at the end of the
drift zone [3]. This consists of a series of electric fields which
repulse the ions back along the flight tube - usually at a slightly
displaced angle (see figure) - resulting in a refocusing of ions with
the same
m/z value on the
reflectron detector.
References:
[1] W.C. Wiley & I.H. MacLaren; The Review of Scientific Instruments,
26; 1955; p1150.
[2] R.J. Cotter; Analytical
Chemistry, 64; 1992,
p1027A.
[3] Mamyrin et al.; Soviet
Physics - JETP, 37; 1973,
p45 and International Journal of
Mass Spectrometry and Ion Processes, 131; 1994, p1.
[4] A tutorial review of TOF-MS was published in 1995: M.
Guilhaus; Journal of Mass
Spectrometry, 30; 1995,
p1519.
©2004 Paul
Gates, University of Bristol
Last updated January 24th 2004